Calibration Procedure - Emerson Micro Motion 3711 Configuration And Use Manual

Gas flow computers
Table of Contents

Advertisement

Calibration
During calibration, the sensor must be completely filled with the calibration fluid, and flow through
the sensor must be at the lowest rate allowed by your application. This is usually accomplished by
closing the shutoff valve downstream from the sensor, then filling the sensor with the appropriate
fluid.
B.4.2

Calibration procedure

To calibrate for density:
1. Click
displayed.
2. Follow the procedure illustrated in Figure B-5.
Note: ROCLINK 800 automatically switches to Auto Scan mode so that the messages will be updated
automatically. This affects only the calibration procedure. When you close the window, Auto Scan will
be returned to its previous setting.
3. If you need to cancel the density calibration process, click the appropriate
button. The Model 3711 will use the pre-existing calibration values.
Figure B-4
Density Calibration panel
200
and open the Density panel. The dialog box shown in Figure B-4 is
Meter > Calibration
Cancel Calibration
®
Micro Motion
Model 3711 Gas Flow Computers

Advertisement

Table of Contents
loading

Table of Contents