Filling The Fixer Tank - Kodak HEALTH SCIENCES X-Omat M43 Installation Instructions Manual

Processor
Table of Contents

Advertisement

INSTALLATION INSTRUCTIONS

Filling the FIXER TANK

Warning
• Wear rubber gloves, safety glasses, and protective clothing when mixing chemicals and filling the TANKS.
• If you are changing solutions in both the DEVELOPER TANK and the FIXER TANK, fill the FIXER TANK
first so that you can thoroughly clean any fixer solution spilled into the DEVELOPER TANK.
• To prevent spills or splashes into other TANKS or onto electrical parts:
– Check that the SIDE ACCESS PANELS are installed.
– Pour the fixer into the TANK at the non-drive side of the PROCESSOR.
– Do not pour fixer onto the RACK.
[1] Check that the fixer DRAIN VALVE is fully closed. Do not overtighten the VALVE.
[2] Mix the required volume of fixer solution in the fixer REPLENISHMENT TANK as recommended by the
instructions packaged with the chemicals.
[3] Install the SPLASH GUARD 1C4019 on the wall between the FIXER TANK and the DEVELOPER TANK.
See the figure for the correct orientation of the SPLASH GUARD.
[4] Carefully pour fixer solution into the non-drive side of the empty FIXER TANK until the level of the solution
in the TANK is at the top of the STEP.
Installing the SPLASH GUARD and Checking the Level of the Fixer Solution
Figure 59
STEP
[5] Very carefully, install the FIXER RACK into the FIXER TANK onto the edge of the SPLASH GUARD. You
can identify the FIXER RACK by the blue WIRE TIE.
(a) Check that the fixer solution did not spill or splash into the DEVELOPER TANK. If necessary, rinse and
dry the DEVELOPER TANK by wiping the TANK with a clean, lint-free cloth.
(b) Check that the RACK is seated firmly in the FIXER TANK.
68
OVERFLOW WEIR
FIXER TANK
SPLASH GUARD
P/N 1C4019
H130_0110HCC
H130_0110HA
September 1995 – 981088