Installing the LIGHTTIGHT GASKET on the PROCESSOR
[54] Cut 3 pieces of the foam GASKET (provided in the THROUGH-THE-WALL KIT) as follows. See the figure.
(a) Cut one GASKET to fit between the 2 MOUNTS for the FEED TRAY.
(b) Cut one GASKET for the area to the right of the MOUNT for the FEED TRAY. Be sure that the GASKET
is long enough to extend 6 mm (
figure below.
(c) Cut one GASKET for the area to the left of the MOUNT for the FEED TRAY. Be sure that the GASKET
is long enough to extend 6 mm (
figure below.
[55] Install the GASKETS to the front of the DEVELOPER TANK WALL as shown below.
[56] Cut a 5.1 cm (2-in.) piece from the foam GASKET for each side of the DEVELOPER TANK WALL.
[57] Install one GASKET on the drive side of the PROCESSOR under the FILM SENSOR WIRES.
[58] Install the other GASKET on the non-drive side.
[59] Advance to the appropriate section as outlined below.
(a) If you are installing an M43A PROCESSOR for 50 Hz operation, advance to the section "Changing the
PROCESSOR to 50 or 60 Hz Operation" on Page 32. Do all the procedures through "Installing the
DRYER RACK" on Page 44.
(b) If you are installing an M43A or Clinic 1 PROCESSOR for 60 Hz operation, advance to the section
"PROCESSOR Setup Options" on Page 35.
Installing the GASKETS to the Front and Sides of the PROCESSOR
Figure 22
FOAM
GASKET
SENSOR
WIRES
WIRE
TIE
981088 – September 1995
1
in.) beyond the edge of the DEVELOPER TANK WALL. See the
4
1
in.) beyond the edge of the DEVELOPER TANK WALL. See the
4
29.5 cm (11 5/8 in.)
17.1 cm (6 3/4 in.)
BRACKET
Installing the PROCESSOR
m m
)
1 9
i n .
m
4
( 3 /
)
6 m
i n .
4
( 1 /
20.3 cm (8 in.)
5.1 cm (2 in.)
DEVELOPER
TANK
SIDEWALL
H130_0116HCA
H130_0116HA