[5] On the 100 CIRCUIT BOARD, move the appropriate JUMPER to its alternate position. Use the chart and
figure below to determine which JUMPER to move.
Table 1 Setup Options
Setup Option
Replenishment (Normal or Flooded) Normal
Developer Temperature
Display Units (Celsius or Fahrenheit)
M43 and M43A Only
Not Used
Water Usage (Continuous or Low)
Standby/Continuous Mode
* If the position of this JUMPER is changed, recalibrate the DEVELOPER TEMPERATURE
DISPLAY. See the Service Manual, Publication No. 981090.
Note
If Low Water Usage Mode is configured, the WASH WATER DRAIN VALVE must be changed. Refer to
"Replacing the Drain Valves," Plumbing Section in the Service Manual.
Changing PROCESSOR Setup Options on the 100 CIRCUIT BOARD
Figure 31
[6] Close the COVER on the ELECTRICAL BOX.
981088 – September 1995
Factory Setting
Celsius*
Continuous
M43:550 and above
M43A: 500 and above
Clinic 1: 600 and above
Low
Remaining serial numbers
Standby
P104
P103
Processor Setup Options
Factory-Set
JUMPER
JUMPER Position
E1
N • • o F
E2
C • • o F
E3
E4
1 • • o 3
1 o • • 3
E5
S • • o C
P105
100 BD.
DIAGNOSTIC
SWITCHES
JUMPERS
H130_0146HCE
H130_0146HA
37