INSTALLATION INSTRUCTIONS
[53] Route the POWER CABLE (provided for the M43A) as necessary for your installation requirements.
(a) Route the CABLE into the MOUNTING STAND in one of the following ways:
• through the notches in the top of either of the side walls
• under the bottom of one of the side walls
(b) It is not recommended to route the POWER CABLE through the darkroom wall. If you must route the
POWER CABLE through the darkroom wall, do the following:
Identify:
• 3 DRAIN HOSES
• HOSE from the main water supply
• POWER CABLE
If the REPLENISHMENT TANKS are located in the MOUNTING STAND, route the HOSES and
POWER CABLE listed above through the:
• closed end of the MOUNTING STAND
• FOAM BLOCK
– you will need to make a hole in the FOAM BLOCK for the POWER CABLE to pass through
• holes in the WOOD FRAME
Positioning the FOAM BLOCK for Installations with the REPLENISHMENT TANKS Located
Figure 21
in the MOUNTING STAND
hole for the
POWER CABLE
DRAIN HOSES
26
MOUNTING STAND
HOSE from the
main water supply
FOAM BLOCK
September 1995 – 981088
H130_0126HCA
H130_0126HA