[12] Insert the INTERLOCK TOOL TL-4740 into the AC INTERLOCK SWITCH S5.
[13] Place a MAGNET onto the DC INTERLOCK SWITCH S6 on the DRYER PLENUM BRACKET.
[14] Check that the main water supply is turned on.
[15] Energize the PROCESSOR by moving the MAIN CIRCUIT BREAKER CB1 to the "1" position.
[16] Allow the PROCESSOR to operate for at least 60 seconds to mix the developer and starter solutions.
Inserting the INTERLOCK TOOL into the INTERLOCK SWITCH
Figure 63
72
INTERLOCK
SWITCH
INTERLOCK
TOOL-4740
H130_0027ACA
H130_0027AA
September 1995 – 981088