Checking for Leaks Behind the
Figure 47
ELECTRICAL BOX
SUPPORT
SHELF
981088 – September 1995
[4] Check that the PROCESSOR is still level when it
[5] Check for leaks at all connections. Tighten HOSE
[6] Pull out the SUPPORT SHELF for the
[7] Pull out the ELECTRICAL BOX.
[8] Using a FLASHLIGHT, check for leaks behind
[9] Push the SUPPORT SHELF and the
[10] Turn on the main water supply to the
Dangerous Voltage
[11] Connect the main power to the PROCESSOR.
ELECTRICAL
[12] Rotate the DRYER TEMPERATURE
BOX
Figure 48
on the DISPLAY PANEL
H130_0151CCA
H130_0151CA
Doing the Water Leak Test
is full of solution. See the procedure on Page 47.
Leveling the PROCESSOR correctly will ensure:
• the correct flow of the processing solutions
• the correct transport of the film
• the correct fit of all ACCESS PANELS
CLAMPS as necessary.
ELECTRICAL BOX from the drive side of the
PROCESSOR.
the ELECTRICAL BOX. Tighten HOSE
CLAMPS, as necessary.
ELECTRICAL BOX to their original positions.
PROCESSOR.
Warning
CONTROL KNOB on the DISPLAY PANEL
fully counterclockwise
Identifying Components
Ready
Wait
Service
DRYER TEMPERATURE
CONTROL KNOB
.
MAIN CIRCUIT
BREAKER CB1
Dryer
Temperature
-
+
H130_0003ACE
H130_0003AA
55