INSTALLATION INSTRUCTIONS
Caution
The empty PROCESSOR weighs approximately 84 kg (185 lb).
[7] Position the PROCESSOR on the MOUNTING STAND. Do not allow the LEVELING FEET to fall into the
holes in the TOP of the MOUNTING STAND.
Using the MOUNTING STAND
Figure 5
holes for
LEVELING FEET
[8] Raise and support the non-feed end of the PROCESSOR to gain access to 2 LEVELING FEET. Use a 5 x 10 cm
(2 x 4 in.) piece of wood that is 1 m (3 ft.) long to support the PROCESSOR.
(a) Position the piece of wood so that:
• it supports the entire width of the PROCESSOR (drive side to non-drive side)
• it is closer to the center of the PROCESSOR than to the back
[9] Remove the NUT from each of the 2 non-feed end LEVELING FEET. Keep the NUTS for later installation.
[10] Remove and discard the LEVELING FOOT on the drive side of the PROCESSOR near the
REPLENISHMENT PUMP.
[11] Install the 7.6 cm (3
[12] Rotate these 2 LEVELING FEET clockwise
ensure that the LEVELING FEET are tight.
[13] Position a SPACER on the TOP of the MOUNTING STAND under each of the 2 LEVELING FEET.
Note
The SPACERS are packed with the MOUNTING STAND.
[14] Lower the raised end of the PROCESSOR by removing the piece of wood supporting it.
[15] Check that the LEVELING FEET are lowered into the SPACERS.
[16] Align the 2 LEVELING FEET and SPACERS with the holes in the TOP of the MOUNTING STAND. Insert
the LEVELING FEET through the holes.
10
1
in.) LEVELING FOOT (packed with the PROCESSOR) in its place. See Figure 6.
2
until they are fully extended. Use a SOCKET or WRENCH to
holes for
LEVELING FEET
H130_0204BCA
September 1995 – 981088
H130_0204BA