3D Reconstruction Upgrade Part 4
7-
Checking C-arm movement for ARCADIS Orbic 3D
63D Reconstruction Upgrade Part 4
Proceed as follows after reconfiguring the basic unit and the monitor trolley.
•
Connect the cables in the zipper hose to the basic unit.
•
Connect the monitor trolley to the line voltage.
•
Release the orbital brake.
•
Manually move the C-arm into both end positions. The movement has to correspond to
"normal" C-arm movement. If not, verify that connector M3.X1 for the brake assembly is
inserted correctly. If yes, the LEDs in both keys light up.
The next step
Continue with ARCADIS Orbic Start-up Instructions SPR2-320.815.02... and Image Qual-
ity Assurance Test SPR2-320.820.02.. .
Perform a complete functional check.