Adjustments / Programs

Configuring Memoskop

Memoskop C-E, keyboard emulation

If no keyboard is included with the present SIREMOBIL Compact with Memoskop C-E,
although a keyboard is required when replacing the Memoskop C-E or when changing the
configuration.
In these cases, the keyboard must be emulated with the assistance of the service PC and
the "TERMINAL.EXE" program. Always use the adapter supplied which is installed on the
keyboard output of the Memoskop C-E. Leave the adapter (item number 31 00 380
G5437) attached to the system.
When replacing the Memoskop C-E, remove the adapter from the Memoskop and attach it
to the new unit.
The Installation / Startup Information MEMOSKOP C-E Keyboard Emulation,
SIREMOBIL Compact up to Serial No. 01416,
Print No. SPR2-130.038.01 and the required
program "Keyboard emulation program terminal.exe"
can be found in the CS Intranet, path:
For Service - SP - SURGERY - SIREMOBIL Compact - CB-DOC - Service - Memoskop.

Configuring the video frequency

After replacing the Memoskop, it will start at a default video frequency (50 Hz). If the mon-
itors were configured for a different frequency, you will have to configure the correct video
frequency without a "visible image".
Proceed with the steps described below:
Invoke Technical Setup on Memoskop (press the CTRL + T keys.)
Press the "2" key and press return. The video frequency menu is selected.
Depending on the required video frequency, press the right cursor key "->".
„->" pressed 1 time:
„->" pressed 2 times:
„->" pressed 3 times:
„->" pressed 4 times:
"<-" = Press the left cursor key.
"->" = Press the right cursor key
Key "->":
Video frequency
Key "<-":
After selecting the desired video frequency, press "Return".
This configures the video frequency you have just selected.
After switching the SIREMOBIL OFF and ON again, the Memoskop C-E will start up at the
correct video frequency.
Siemens AG
Medical Engineering
60 Hz video frequency
100 Hz video frequency
120 Hz video frequency
50 Hz video frequency.