View Test Data - Emerson Micro Motion 2700 Configuration And Use Manual

With foundation fieldbus
Hide thumbs Also See for Micro Motion 2700:
Table of Contents

Advertisement

Measurement support
8.2.4

View test data

You can view the results of the current test. You can also view results from previous tests.
Important
You can view previous test results and see detailed test reports only if SMV is licensed.
The transmitter stores the following information about the previous twenty SMV tests:
In addition, ProLink III provides a detailed test reporting and analysis framework. This
information is stored on the PC where ProLink III is installed for tests that were run only on
that PC. It includes:
You can use ProLink III to run a test that displays a test result chart and a test report at the
completion of the test.
View test result data using the display
1.
2.
94
Powered-on hours at the time of the test.
Test result (Pass, Fail, Abort).
Abort code, if applicable.
Timestamp from the PC clock
Current flowmeter identification data
Current flow and density configuration parameters
Current zero values
Current process values for mass flow rate, volume flow rate, density, temperature,
and external pressure
Customer and test descriptions (if entered by the user)
If you have just run a test, results are displayed automatically at the end of the test.
If SMV is licensed, and you want to view results from previous tests:
a. Navigate to the Smart Meter Verification menu.
Figure 8-3:
SMV – Top-level menu
RUN VERFY
Scroll
Select
Micro Motion Model 2700 Transmitters with FOUNDATION Fieldbus
Scroll and Select simultaneously
for 4 seconds
Scroll
ENTER METER VERFY
Select
RESULTS READ
SCHEDULE VERFY
Scroll
Select
EXIT
Scroll
Select
Scroll
Select