Procedure
Set Volume Flow Type to Gas Standard Volume.
4.3.2

Configure Standard Density of Gas

Display
ProLink III
Field Communicator Configure > Manual Setup > Measurements > GSV > Gas Ref Density
Fielbus host
Overview
The Standard Density of Gas value is the gas density at standard reference conditions. Use
it to convert the measured mass flow data to volume flow at reference conditions.
Prerequisites
Ensure that Density Measurement Unit is set to the measurement unit you want to use for
Standard Density of Gas.
Procedure
Set Standard Density of Gas to standard reference conditions.
4.3.3

Configure Gas Standard Volume Flow Unit

Display
ProLink III
Field Communicator Configure > Manual Setup > Measurements > GSV > GSV Flow Unit
Fieldbus host
Overview
Gas Standard Volume Flow Unit specifies the unit of measure that will be displayed for the
gas standard volume flow. The measurement unit used for the gas volume total and the
gas volume inventory is derived from this unit.
Prerequisites
Before you configure Gas Standard Volume Flow Unit, be sure that Volume Flow Type is set
to Gas Standard Volume.
Procedure
Set Gas Standard Volume Flow Unit to the unit you want to use.
Configuration and Use Manual
Not available
Device Tools > Configuration > Process Measurement > Flow
Measurement TB > GSV_Gas_Dens (OD Index 62)
OFF-LINE MAINT > OFF-LINE CONFG > UNITS > GSV
Device Tools > Configuration > Process Measurement > Flow
Measurement TB > SNS_GSV_FlowUnits (OD Index 67)
Configure process measurement
27