Transmitter status reported by LED

Table 2-1:
Transmitter status reported by status LED
LED state
Description
Solid green
No alerts are active.
Flashing green (if ena-
Unacknowledged corrected condition (no
bled)
alert)
Solid yellow
One or more low-severity alerts are active.
A low severity alarm can mean one or more
process variables is at a set output level (i.e.
simulation or two phase timeout).
Flashing yellow (if ena-
Calibration in progress.
bled)
One or more low-severity alerts are active and
have not been acknowledged.
Solid red
One or more high-severity alerts are active.
Flashing red (if ena-
One or more high-severity alerts are active
bled)
and have not been acknowledged.
If Status LED Blinking is disabled, all LEDs will show a solid color rather than flashing.
6
Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and
checks for error conditions. During the power-up sequence, Alert A009 is active.
This alert should clear automatically when the power-up sequence is complete.
Check the status LED on the transmitter.
View and acknowledge status alerts
Micro Motion Model 2700 Transmitters with FOUNDATION Fieldbus
Recommendation
Continue with configuration or process meas-
urement.
Continue with configuration or process meas-
urement. Acknowledge the alert if you choose.
A low-severity alert condition does not affect
measurement accuracy or output behavior.
You can continue with configuration or proc-
ess measurement, but Micro Motion still rec-
ommends identifying and resolving the alert
condition.
A low-severity alert condition does not affect
measurement accuracy or output behavior.
You can continue with configuration or proc-
ess measurement, but Micro Motion still rec-
ommends identifying and resolving the alert
condition.
A high-severity alert condition affects meas-
urement accuracy and output behavior. Re-
solve the alert condition before continuing.
A high-severity alert condition affects meas-
urement accuracy and output behavior. Re-
solve the alert condition before continuing.
Acknowledge the alert if you choose.