Run Smv - Emerson Micro Motion 2700 Configuration And Use Manual

With foundation fieldbus
Hide thumbs Also See for Micro Motion 2700:
Table of Contents

Advertisement

SMV has an output mode called Continuous Measurement that allows the transmitter to keep
measuring while the test is in progress. If you choose to run the test in Last Measured Value
or Fault modes instead, the transmitter outputs will be held constant for the two minute
duration of the test. If control loops depend on transmitter outputs, take appropriate
action.
Avoid process instability during the test. If conditions are too unstable, the SMV test will
abort. To maximize process stability:
Tips
8.2.3

Run SMV

Run an SMV test using the display
1.
2.
3.
Configuration and Use Manual
Maintain a constant fluid pressure and temperature.
Avoid changes to fluid composition, e.g., two-phase flow or settling.
Maintain a constant flow rate.
The SMV test runs best when flow is stopped through the sensor.
SMV is not affected by any configured transmitter parameters for flow, density, or
temperature.
Navigate to the Smart Meter Verification menu.
Figure 8-1:
SMV – Top-level menu
RUN VERFY
Scroll
Select
Choose Run Verify.
Choose Outputs and select the desired output behavior.
Option
Description
Continue
During the test, all outputs will continue to report their assigned process var-
Measr
iable. The test will run for approximately 90 seconds.
Fault
During the test, all outputs will go to their configured fault action. The test
will run for approximately 140 seconds.
Scroll and Select simultaneously
for 4 seconds
Scroll
ENTER METER VERFY
Select
RESULTS READ
Scroll
SCHEDULE VERFY
Select
Select
Measurement support
Scroll
EXIT
Scroll
Select
91