Troubleshooting
Alert num-
ber
Alert title
A131
Meter Verification in
Progress: Outputs to
Last Measured Value
A132
Sensor Simulation
Active
A133
EEPROM Error (Dis-
play)
N/A
Density FD Calibra-
tion in Progress
N/A
Density D1 Calibra-
tion in Progress
N/A
Density D2 Calibra-
tion in Progress
N/A
Density D3 Calibra-
tion in Progress
N/A
Density D4 Calibra-
tion in Progress
N/A
Zero Calibration in
Progress
N/A
Reverse Flow
9.2
Flow measurement problems
Problem
Possible causes
Non-zero flow reading
•
at no-flow conditions
or at zero offset
•
•
116
Possible cause
A meter verification test is in pro-
gress, with outputs set to Last Meas-
ured Value.
Sensor simulation is enabled.
There is a memory error in the dis-
play module.
A flowing density calibration is in
progress.
A D1 density calibration is in pro-
gress.
A D2 density calibration is in pro-
gress.
A D3 density calibration is in pro-
gress.
A D4 density calibration is in pro-
gress.
A zero calibration is in progress.
Flow through the device is in the re-
verse direction (against the flow ar-
row).
Misaligned piping (especially in new instal-
lations)
Open or leaking valve
Incorrect sensor zero
Micro Motion Model 2700 Transmitters with FOUNDATION Fieldbus
Recommended actions
•
Allow the procedure to complete.
•
Disable sensor simulation.
•
Cycle power to the meter.
•
Replace the display module.
•
Contact customer support.
•
No action required.
•
No action required.
•
No action required.
•
No action required.
•
No action required.
•
No action required.
•
No action required.
Recommended actions
•
If the reading is not excessively high, re-
view the live zero. You may need to restore
the factory zero.
•
Check for open or leaking valves or seals.
•
Check for mounting stress on the sensor
(e.g., sensor being used to support piping,
misaligned piping).
•
Contact customer support.