Configure process measurement
4.2
Configure volume flow measurement for liquid
applications
The volume flow measurement parameters control how liquid volume flow is measured
and reported.
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time.
Choose one or the other.
4.2.1
Configure Volume Flow Type for liquid applications
Display
ProLink III
Field Communicator Configure > Manual Setup > Measurements > GSV > Volume Flow Type > Liquid
Fieldbus host
Overview
Volume Flow Type controls whether liquid or gas standard volume flow measurement will
be used.
Restriction
Gas standard volume measurement is incompatible with some applications. Set Volume Flow Type
to Liquid if you are using any of the following applications:
•
•
Procedure
Set Volume Flow Type to Liquid.
4.2.2
Configure Volume Flow Measurement Unit for liquid
applications
Display
ProLink III
Field Communicator Configure > Manual Setup > Measurements > Flow > Volume Flow Unit
Fieldbus host
22
Not available
Device Tools > Configuration > Process Measurement > Flow
Measurement TB > SNS_EnableGSV (OD Index 66)
Petroleum measurement
Concentration measurement
OFF-LINE MAINT > OFF-LINE CONFG > UNITS > VOL
Device Tools > Configuration > Process Measurement > Flow
Measurement TB > VOLUME_FLOW_UNITS (OD Index 25)
Micro Motion Model 2700 Transmitters with FOUNDATION Fieldbus