Subsystem Syntax
:COUNt
Comments
78 Digitizer Command Reference
The SAMPle command subsystem sets the number of samples to be taken for each
trigger. It also sets the number of samples to be taken prior to the trigger and the
source of the sample signal and its slope. When the sample source is TIMer, you can
also set the sample interval.

SAMPle

:COUNt < count > | MIN | MAX
:COUNt? [MIN | MAX]
[:IMMediate]
:PRETrigger:COUNt < count > | MIN | MAX
:PRETrigger:COUNt? [MIN | MAX]
:SLOPe POS | 1 | NEG | 0
:SLOPe?
:SOURce HOLD | TIMer | TTLT0-7 | EXT
:SOURce?
:TIMer < interval > | MIN | MAX
:TIMer? [MIN | MAX]
SAMPle:COUNt <count> | MIN | MAX command sets the number of total
samples which includes the pre-trigger and post-trigger samples. The number of
samples set is common to all channels. You cannot have two or more channels with
different sample settings.
The total number of readings is limited to at most 16,777,216 for the 4-channel
SAMPle