ROSCillator:EXTernal:FREQuency
Parameters
Comments
ROSCillator:EXTernal:FREQuency?
74 Digitizer Command Reference
NOTE: Use the [SENSe:]VOLTage[:DC]:RESolution?
command to obtain the resolution value.
Data is returned as real numbers when the data format is set to REAL (see the
FORMat[:DATA] REAL command). The data is returned in voltage units
and no scaling conversion is required as with the PACKed format.
Both PACKed and REAL formats return data preceded by the IEEE-488.2
definite length arbitrary block header. The header is as follows:
# <num_digits> <num_bytes>
# signifies a block transfer
<num_digits> is a single digit (1 through 9) specifying how many digits
(ASCII characters) are in the <num_bytes> descriptor which follows
<num_bytes> is the number of data bytes which immediately follow the
<num_bytes> field
[SENSe:]ROSCillator:EXTernal:FREQuency <freq> is used to specify the
externally supplied timebase frequency. This command is not required unless
ROSCillator:SOURce is EXTernal. The default timebase is the INTernal timebase.
Parameter
Name
freq
The frequency parameter value is used to calculate sample periods when the
sample source is set to TIMer.
The sample period must be at least 1.250E-6 seconds (800 kHz), and must be
an integral multiple of the timebase period (this is 1.0E-7 seconds when the
timebase source is INTernal). Period values will be rounded to the nearest
period the instrument can obtain.
Executable when initiated: NO
Coupled command: NO
Reset (*RST) Condition: frequency = 10.0 MHz
[SENSe:]ROSCillator:EXTernal:FREQuency? queries the external frequency.
Parameter
Range of
Type
numeric
9.9E3 to 30E6
Default
Values
Units
Hz