:DAC:SOURce
Parameters
Comments
:DAC:SOURce:RAMP
Parameters
Comments
DIAGnostic:DAC:SOURce <voltage> outputs the specified voltage from the
internal calibration source DAC onto the cal pins of the front panel connector. This
command is a factory diagnostic routine.
Parameter
Name
voltage
The channel's input relay remains open until it is closed by an INPut:STATe
ON command or by a reset of the instrument.
Executable when initiated: No
Coupled Command: No
Reset (*RST) Condition: DAC output is set to 0V
DIAGnostic:DAC:SOURce:RAMP <count> outputs a ramp of DAC values
from 0 to 4095 with the DAC code changing about every 100 µS. This command is
a factory diagnostic routine.
Parameter
Name
count
The "count" variable specifies how many ramps to output. The timing is such
that about 2.3257 full ramps are output each second.
A count of 139 will output ramps for just under 60 seconds.
The signal will be output onto the cal pins on the front panel D-subminiature
connector.
Executable when initiated: No
Coupled Command: No
Reset (*RST) Condition: DAC output is set to 0V
Parameter
Range of
Type
numeric
-15.0 to +15.0
Parameter
Range of
Type
numeric
1 through 255
Digitizer Command Reference
Default
Values
Units
none
Default
Values
Units
none