:PRETrigger:COUNt?
[:IMMediate]
:SLOPe
Comments
:SLOPe?
80 Digitizer Command Reference
This count specifies the portion of the total SAMPle:COUNt that will be
sampled prior to the trigger.
A trigger is ignored if it occurs before the pretrigger count is met.
If the specified number of pretrigger samples (<count>) have been taken and a
trigger has not yet occurred, the digitizer continues to sample the input signal.
The digitizer retains the most recent pretrigger samples specified by the
number "<count>" when the trigger does occur.
Executable when initiated: No
Coupled command: No
Reset (*RST) condition: 0 pretriggers
SAMPle:PRETrigger:COUNt? [MIN | MAX] command returns the number of
pretrigger samples each channel will make prior to each trigger. The number of
pretriggers returned is common to all channels.
SAMPle[:IMMediate] command is generally used only when the sample source is
HOLD to take a single reading when the digitizer is in the wait-for-sample state.
SAMPle:SLOPe POS | 1 | NEG | 0 command sets the slope of the sample signal
(the active edge, rising or falling, of the sample signal). The slope setting is common
to all channels.
This command is effective only when the sample source is EXTernal. The
slope is set but will be ignored if the sample source is a source other than
EXTernal.
Executable when initiated: No
Coupled command: No
Reset (*RST) condition: POSitive (1)
SAMPle:SLOPe? command queries the present setting of the slope of the sample
signal. The sample slope is effective only when the sample source is EXTernal.

Advertisement