HP E1563A User Manual page 97

Table of Contents

Advertisement

:SLOPe[<n>]?
Parameters
Note
:SOURce[<n>]
Note
Parameters
Comments
Reset (*RST) condition: SLOPe1 = POSitive; SLOPe2 = POSitive
TRIGger:SLOPe[<n>]? queries the present setting for the slope of the trigger
signal for the trigger source (1 or 2) specified. Trigger slope for source number 1 is
returned if "n" is not designated.
Parameter
Name
n
Trigger slope applies only for TRIG:LEVel when the trigger source is INTernal or
when EXTernal trigger sources are specified.
Returns either "POS" or "NEG".
TRIGger:SOURce[<n>]
TTLT0-7 sets the source of the trigger for all channels or can disable the trigger source.
command defaults to trigger source number 1 if "n" is not designated.
Two trigger sources are allowed, TRIG:SOUR1 and TRIG:SOUR2, which are common to
ALL channels on the E1563A and E1564A. SOUR1 is not associated only with channel 1
and SOUR2 is not associated only with channel 2.
Parameter
Name
n
The TRIGger:SOURce command only selects the trigger source. You must
use the INITiate command to place the digitizer in the wait-for-trigger state.
TRIGger:SOURce EXT uses the digitizer's front panel "Trig" pin in the D
sub-miniature connector as the trigger source. The digitizer triggers on the
falling (negative-going) edge of a ±5V TTL input signal (maximum input is
+5V peak to the front panel D sub-miniature connector "Trig" pin).
TRIGger:IMMediate causes a trigger to occur immediately provided the
Parameter
Range of
Type
numeric
BUS | EXTernal | HOLD | IMMediate | INTernal1-4 |
Parameter
Range of
Type
numeric
Digitizer Command Reference
Default
Values
Units
1 or 2
None
Default
Values
Units
1 or 2
None