HP E1563A User Manual page 77

Table of Contents

Advertisement

Parameters
Comments
VOLTage[< channel >][:DC]:RANGe?
VOLTage[< channel >][:DC]:RESolution?
The crossover points for range changes are as follows:
Voltage Range
Resolution
0.0625
.000007629
0.2500
.000030518
1.0000
.000122070
4.0000
.000488281
16.0000
.007812500
64.0000
.007812500
256.0000
.03125
Executable when initiated: No
Coupled command: YES; TRIGger:LEVel may be affected if one of the
Reset condition: range is set to 256 for all channels
[SENSe:]VOLTage[<channel>][:DC]:RANGe? queries the specified channel for
its present range setting. The command defaults to channel 1 if no channel is
specified.
[SENSe:]VOLTage[<channel>][:DC]:RESolution? queries the specified channel
for its present resolution setting. Resolution versus range setting is shown in the
VOLTage[:DC]:RANGe command. The command defaults to channel 1 if no
channel is specified.
levels is the trigger event on the channel that had the
range change. The level set for CALCulate:LIMit
:LOWer (and :UPPer) will be modified to be the same
percent of full range. This will generate a different
voltage value for the limit level.
Digitizer Command Reference
77