*OPC?
Comments
*RST
Comments
•
Coupled command: No
•
Related commands: *OPC?, *WAI
•
Reset (*RST) condition: none
*OPC? causes the HP E1563A and E1564A Digitizers to wait for all pending
operations to complete. A single ASCII "1" is then placed in the output queue.
•
The INIT command is considered complete when the measurement is started.
*OPC? will return "1" once INIT is processed and measurements start but the
instrument may not be finished taking all readings initiated.
•
Executable when initiated: Yes
•
Coupled command: No
•
Related commands: *OPC?, *WAI
•
Reset (*RST) condition: none
*RST resets the HP E1563A and E1564A Digitizers as follows:
•
Sets all commands to their *RST state.
•
Aborts a calibration (CAL:STATe ON), resetting the CAL:STATe to OFF.
•
Aborts all pending operations.
*RST does not affect:
•
The output queue
•
The Service Request and Standard Event Status Enable Registers
•
The enable unmasks for the Questionable Signal Registers
•
Calibration data
•
Executable when initiated: Yes
•
Coupled command: No
•
Reset (*RST) condition: none
Digitizer Command Reference