-160 to -168
140 HP E1563A and E1564A Digitizer Error
Example: SAMP:COUN
-112
Program mnemonic too long
A command header was received which contained more than the maximum
12 characters allowed. Example: SAMPLE:PRETRIGGER:COUNT 10
-113
Undefined header
A command was received that is not valid for this digitizer. You may have
misspelled the command or it may not be a valid command. If you are using
the short form of the command, remember that it may contain up to four
letters. Example: TRIGG:LEV 1.2
-121
Invalid character in number
An invalid character was found in the number specified for a parameter
value. Example: STAT:QUES:ENAB #B01010102
-123
Numeric overflow
A numeric parameter was found whose exponent was larger than 32,000.
Example: SAMP:COUN 1E34000
-124
Too many digits
A numeric parameter was found whose mantissa contained more than 255
digits, excluding leading zeros.
-128
Numeric data not allowed
A numeric parameter was found but a character string was expected. Check
the list of parameters to verify you have used a correct parameter type.
Example: TRIG:SOUR 2 EXT (should be TRIG:SOUR2 EXT)
-138
Suffix not allowed
A suffix was received following a numeric parameter which does not accept
a suffix. Example: SAMP:COUN 1 SEC (
-148
Character data not allowed
A character string was received but a numeric parameter was expected.
Check the list of parameters to verify that you have used a valid parameter
type. Example: CAL:VAL XYZ
-158
String data not allowed
A character string was received but is not allowed for the command. Check
the list of parameters to verify that you have used a valid parameter type.
Example: CALC:LIM:LOW:STAT 'ON'
Block data errors
The digitizer does accept block data.
is not a valid suffix).