HP E1563A/E1564A Accuracy Specifications
Zero Offset
(with filter OFF)
Specifi-
Range
2
cation
20 µV
0.0625V
78 µV
0.25V
300 µV
1V
4V
1.2 mV
16V
21 mV
64V
28 mV
256V
79 mV
1
Valid within the range of 0°C to 55°C. A zero offset calibration for all channels must be performed
if the instrument experiences a temperature <0°C or >55°C for these specifications to remain valid.
2
Specification is valid when tested at a temperature within ±5°C of the calibration temperature.
3
Amount of error that must be added for each °C outside of ±5°C of the calibration temperature.
Integral Non-linearity Specification
All ranges: 2.5 LSB
General Specifications
For indoor use, pollution degree 2
Operating altitude: 3000 meters or mainframe altitude specification, whichever is lower
Operating temperature: 0°C to 55°C
Relative humidity: up to 80% at 31°C, decreasing to 50% at 40°C
112 HP E1563A and E1564A Digitizer
1
Zero Offset
(with filter ON)
Temperature
Specifi-
3
Coefficient
cation
1.9 µV/°C
28 µV
6 µV/°C
110 µV
15 µV/°C
430 µV
60 µV/°C
1.7 mV
1.3 mV/°C
21 mV
1.65 mV/°C
34 mV
4.28 mV/°C
110 mV
1
Temperature
Specifi-
2
3
Coefficient
cation
4.3 µV/°C
0.034%
16 µV/°C
0.034%
63 µV/°C
0.034%
251 µV/°C
0.034%
1.63 mV/°C
0.034%
4.24 mV/°C
0.034%
16.2 mV/°C
0.034%
Gain
(% of reading)
(3 sigma)
Temperature
2
3
Coefficient
0.0061%/°C
0.0061%/°C
0.0061%/°C
0.0061%/°C
0.0061%/°C
0.0061%/°C
0.0061%/°C
Noise
Specifi-
cation
57 µV
180 µV
720 µV
2.88 mV
14.7 mV
48 mV
189 mV

Advertisement