Subsystem Syntax
:DAC:OFFSet
Parameters
Comments
:DAC:OFFSet:RAMP
The DIAGnostic subsystem contains several commands which were developed to
test the instrument at the factory. Some of these commands may prove useful for
isolating problems or for use in special applications.

DIAGnostic

:DAC:OFFSet[< channel >] < voltage >
:DAC:OFFSet[< channel >]:RAMP < count >
:DAC:GAIN[< channel >] < value >
:DAC:SOURce < voltage >
:DAC:SOURce:RAMP < count >
:INTerrupt:LINE 0 | 1 | 2 | 3 | 4 | 5 | 6 | 7
:INTerrupt:LINE?
:MEMory:SIZE < size >
:MEMory:SIZE?
:PEEK? < reg_number >
:POKE? < reg_number >,< data >
:SHORt[< channel >] ON | 1 | OFF | 0
:SHORt[< channel >]?
:STATus?
DIAGnostic:DAC:OFFSet
value to the calibration offset DAC of the specified channel when the
command is sent. This offset voltage value is not used unless a
is sent to the calibration gain DAC. This command is a factory diagnostic routine.
Parameter
Name
voltage
Executable when initiated: No
Coupled Command: No
Reset (*RST) Condition: None
DIAGnostic:DAC:OFFSet
channel, a ramp of DAC values from 0 to 255 with the DAC code changing
approximately every 100 µS. This command is a factory diagnostic routine.
[< channel >]
<voltage> writes the specified voltage
Parameter
Range of
Type
numeric
-2.5 to +2.5
:RAMP <count> outputs to the specified
[< channel >]
Digitizer Command Reference
DIAGnostic
DAC:GAIN
DAC:GAIN < value >
Default
Values
Units
none
55