:LIMit:UPPer:DATA
Parameters
Comments
:LIMit:UPPer:DATA?
46 Digitizer Command Reference
CALCulate[<channel>]:LIMit:UPPer:DATA <value> | MIN | MAX sets the
upper limit value you want to test against. The :LIMit:FAIL? command will return
a "1" following the measurement (and prior to the next measurement) if the input
signal rose above the specified upper limit value and LIM:UPP:STATe is ON. A "0"
is returned if the limit was not exceeded.
Parameter
Name
value
•
The maximum allowed <value> depends on the range setting. An error will
occur if you try to set a level that exceeds the range setting.
•
Changing the range after setting the limit value will change the limit value.
The percent of full scale is kept constant.
•
Allowable maximum values for the upper limit by range and the associated
resolution is given below:
Range
Maximum value
0.0625
0.062011719
0.250
0.248046875
1.00
0.992187500
4.00
3.968750
16.00
15.8750
64.00
63.50
256.00
254.00
•
Executable when initiated: No
•
Coupled Command: YES; Range changes will change the value; the
•
Reset (*RST) Condition: +252 volts
CALCulate[<channel>]:LIMit:UPPer:DATA? [MIN | MAX] queries the upper
limit value set for the specified channel.
Parameter
Range of
Type
numeric
-254 to +252
Resolution
0.000488281
0.001953125
0.00781250
0.031250
0.1250
0.500
2.0
percent of full scale of the range will be kept constant.
For example, on the 4 volt range, with a 2V limit, a
range change to 16V will set a new limit of 8V.
Default
Values
Units
volts