:LIMit:UPPer:DATA
Parameters
Comments
:LIMit:UPPer:DATA?
46 Digitizer Command Reference
CALCulate[<channel>]:LIMit:UPPer:DATA <value> | MIN | MAX sets the
upper limit value you want to test against. The :LIMit:FAIL? command will return
a "1" following the measurement (and prior to the next measurement) if the input
signal rose above the specified upper limit value and LIM:UPP:STATe is ON. A "0"
is returned if the limit was not exceeded.
Parameter
Name
value
The maximum allowed <value> depends on the range setting. An error will
occur if you try to set a level that exceeds the range setting.
Changing the range after setting the limit value will change the limit value.
The percent of full scale is kept constant.
Allowable maximum values for the upper limit by range and the associated
resolution is given below:
Range
Maximum value
0.0625
0.062011719
0.250
0.248046875
1.00
0.992187500
4.00
3.968750
16.00
15.8750
64.00
63.50
256.00
254.00
Executable when initiated: No
Coupled Command: YES; Range changes will change the value; the
Reset (*RST) Condition: +252 volts
CALCulate[<channel>]:LIMit:UPPer:DATA? [MIN | MAX] queries the upper
limit value set for the specified channel.
Parameter
Range of
Type
numeric
-254 to +252
Resolution
0.000488281
0.001953125
0.00781250
0.031250
0.1250
0.500
2.0
percent of full scale of the range will be kept constant.
For example, on the 4 volt range, with a 2V limit, a
range change to 16V will set a new limit of 8V.
Default
Values
Units
volts