Parameters
Comments
SWEep:POINts?
SWEep:OFFSet:POINts
Parameters
Note
Comments
SWEep:OFFSet:POINts?
VOLTage[< channel >][:DC]:RANGe
76 Digitizer Command Reference
Parameter
Name
<count>
**(memory size in bytes) / (number of channels * 2) = 128M/4 or 128M/8 (MAX)
This command is the same as SAMPle:COUNt and is included for SCPI
compatibility.
[SENSe:]SWEep:POINts? [MIN | MAX] command returns the sweep points.
[SENSe:]SWEep:OFFSet:POINts <count> | MIN | MAX command sets the
number sweep offset points.
<count> must be a negative number.
<count> must be a negative number.
This command is the same as
<count> is negative here, whereas, it is positive for pretrigger count) and is
included for SCPI compatibility.
[SENSe:]SWEep:OFFSet:POINts? [MIN | MAX] command returns the sweep
offset points.
[SENSe:]VOLTage[<channel>][:DC]:RANGe <range> command is used to
change the range on the specified channel. There are seven different ranges. If the
range specified falls between two of the instrument's ranges, the range is set to the
next higher range setting. The command defaults to channel 1 if no channel is
specified.
Parameter
Range of
Type
numeric
1 to 32M** (E1563A)
1 to 16M** (E1564A)
SAMPle:PRETrigger:COUNt
Default
Values
Units
N/A
(except the sign on

Advertisement