:INTerrupt:LINE
Note
Comments
:INTerrupt:LINE?
:MEMory:SIZE
Note
Parameters
Comments
58 Digitizer Command Reference
DIAGnostic:INTerrupt:LINE 0 | 1 | 2 | 3 | 4 | 5 | 6 | 7 sets the interrupt line to be
used. Specifying the "0" parameter disables interrupts.
The STATus subsystem will not work if interrupts are disabled (STATus:
OPEReration and STATus:QUEStionable commands). Use the DIAG:STATus?
command when you disable interrupts.
The "0" parameter disables all interrupts.
Power-on default setting is interrupt line "1".
Executable when initiated: No
Coupled Command: No
Reset (*RST) Condition: Interrupt line setting is unchanged.
DIAGnostic:INTerrupt:LINE? queries the interrupt line setting. Returns a
number "0" through "7" to indicate interrupt line 1 through 7. A "0" returned
indicates interrupts are disabled. See Note in preceding DIAG:INTerrupt:LINE.
DIAGnostic:MEMory:SIZE <size> sets the memory size value in calibration
memory. Your module comes standard with 4 Mbytes of RAM. You can replace
this with PC SIMM modules of up to 128 Mbytes. See Chapter 1 for the procedure
for adding RAM to your module.
This command is required and used only when you change the size of RAM on the
module. You then use this command to set the new memory size value in
calibration memory.
Parameter
Name
size
The <size> parameter will accept a value in excess of the industry notation value of
Parameter
Range of
Type
numeric
4E6, 8E6, 16E6, 32E6,
64E6 and 128E6
Default
Values
Units
none