Analyzer Slider - Nikon Eclipse LV100DA-U Instructions Manual

Universal design microscope udn
Table of Contents

Advertisement

16

Analyzer Slider

The polarization microscopy under the episcopic illumination
can be performed when the analyzer slider is used with the
polarizer slider for the episcopic illumination, and the
polarization microscopy under the episcopic illumination can
be performed when the analyzer slider is used with the
polarizer for the diascopic illumination. Additionally, the
sensitive color polarization microscopy under the episcopic
illumination can be performed when the lambda plate is
attached to the configuration for the polarization microscopy
under the episcopic illumination, and the sensitive color
polarization microscopy under the diascopic illumination can
be performed when the lambda plate is attached to the
configuration for the polarization microscopy under the
diascopic illumination.
The DIC microscopy under the episcopic illumination can be performed when the analyzer slider is
used with the LV-UPO polarizer slider for the episcopic illumination equipped with the1/4 lambda
plate and the objective DIC prism, and the sensitive color DIC microscopy under the episcopic
illumination can be performed when the lambda plate slider is attached to the configuration of the
DIC microscopy under the episcopic illumination.
The DIC microscopy under the diascopic illumination can be performed when the universal
condenser, the analyzer slider, D-DP rotatable polarizer for the diascopic illumination equipped
with the 1/4 lambda plate, the condenser DIC prism, and the objective DIC prism are attached to the
microscope. Additionally, the sensitive color DIC microscopy under the diascopic illumination can
be performed when the lambda plate slider is attached to the configuration for the DIC microscopy
under the diascopic illumination.
Placing the analyzer into the optical path
• Attaching the analyzer slider: Remove the horizontally oriented cover on the right side of the
illuminator. And then, insert the analyzer slider into the horizontal slot with its mark facing up.
(See Page 83.)
• Placing the element into the optical path: When the analyzer slider is pushed into the first
click-stop position, the empty hole is placed into the optical path. And when the analyzer slider is
pushed into the second click-stop position, the analyzer is placed into the optical path.
* See the arrow in the figure below for the analyzer orientation.
Removing the analyzer from the optical path
If the analyzer is placed in the optical path and you
wish to remove the analyzer from the optical path,
pull the analyzer slider toward the right to the first
click-stop position. (The empty hole will be placed
into the optical path.)
Analyzer orientation
The orientation of the analyzer is the forward and
backward of the microscope as shown in the right
figure.
60
First clickstop position
Second clickstop position
Analyzer
P
F. ST O
BF
DF FL1 FL2
FL1
FL2
UEP I2A
USB
Analyzer slider