Attaching the Double Light Source Adapter

To perform the microscopy under the episcopic illumination, attach the LVUEPI2-DLS double light
source adapter. The standard halogen lamp and the external light source can be attached together
and switched with the light sources lit.
CAUTION
1 Check that the microscope and the light source are
turned off.
2 Insert the double light source adapter into the
mounting part of the LV-UEPI2A as far as it goes
and tighten the clamp screw securely with the
hexagonal screwdriver provided with the
microscope.
Attach the adapter with the clamp screw located on
the top and the slide knob on the right.
To tighten the clamp screw, insert the hexagonal
screwdriver through the cooling slit into the
adapter and tighten the clamp screw inside.
3 Attach the lamp house to the lamp house mounting
part of the double light source adapter as far as it
goes and tighten the clamp screw securely with the
hexagonal screwdriver.
For the connection of the lamp cable, see Page 85,
"Attaching the Lamp House and Replacing
Lamps."
4 Insert the tip of the light guide fiber into the light
guide fiber connecting part of the product as far as
it goes and tighten the clamp screw securely.
For the procedure of connecting the external light
source and the light guide fiber, see the instruction
manual for the light source.
92
Before using the double light source adapter, make sure that the clamp screws on each
part are tightened securely.
Clamp screw
Light source
mounting part
A. ST O
P
P
F. ST O
Lamp house mounting part
Clamp screw
LA MP
DC 12V
50W
LV-LH50PC
Fiber clamp screw
Light guide
fiber
To the external
light source
LA MP
50W
DC 12V
Connecting part
Slide knob