Analyzer Slider; Attaching The Analyzer Slider; Placing/Removing The Analyzer In/From The Optical Path - Nikon ECLIPSE L200ND / ECLIPSE L200N Instructions Manual

Ic inspection microscope
Table of Contents

Advertisement

3.12

Analyzer slider

Use together with the polarizer slider for episcopic bright-field simplified polarization
microscopy.
Use together with the polarizer and the DIC sliders for episcopic DIC microscopy.
Empty hole

3.12.1 Attaching the analyzer slider

Remove the cap near the indication "ANALYZER" at the right side of the microscope arm.
Insert the analyzer slider with its nameplate facing up. Slide the slider in and out of the
microscope to find two click-stop positions on the slider.

3.12.2 Placing/removing the analyzer in/from the optical path

• Placing the analyzer in the optical path
Push in the slider to the second click-stop position to place the analyzer in the optical
path.
Since the orientation of the analyzer is already adjusted at the factory, you can easily get
the crossed Nicols position by just placing the polarizer and the analyzer in the optical
path.
• Removing the analyzer from the optical path
Pull out the slider to the first click-stop position to remove the analyzer from the optical
path. (The empty hole is now placed in the optical path.)
3 Operation of Each Part
Figure-28 Analyzer slider
Figure-29 Analyzer
24
IC Inspection Microscope ECLIPSE L200ND / L200N Instructions
Analyzer
Analyzer slider
Push in:
The analyzer is in
the optical path.