5 Rotate the rotation ring of the polarizer slider to adjust the contrast. (See Page 57.)
The polarizer slider is equipped with the 1/4 lambda plate so that the contrast can be adjusted by
adjusting the orientation of the polarizer.
6 To perform the sensitive color microscopy, push in the lambda plate slider to place the lambda
plate into the optical path. (See Page 62.)
7 Adjust the EPI brightness switch on the operation panel to adjust the brightness of the episcopic
illumination. (See Page 42.)
8 Adjust the brightness with ND filters. (See Page 44.)
Return to the bright-field microscopy under the
4
episcopic illumination.
1 Pull out the analyzer slider to remove the analyzer
from the optical path. (See Page 60.)
2 Pull out the polarizer slider to remove the polarizer
from the optical path. (See Page 57.)
3 Pull out the DIC slider to remove the DIC prism
from the optical path. (See Page 64.)
4 Pull out the lambda plate slider to remove the
lambda plate from the optical path.
(See Page 62.)
5 Operate the EPI brightness switch on the
operation panel to adjust the brightness of the
episcopic illumination. (See Page 42.)
6 Adjust the brightness with ND filters.
(See Page 44.)
II. Microscopy Method
1 2
F. ST OP
BF
DF FL1 FL2
FL1
FL2
4
6
3
0.8
0.7
Achr N.A
0.6
0.5 0.4
= 0.9
JAPAN
0.3 0.2 0.1
Operation
panel
NCB
ND8
F.S.
5
EPI
DIA
A.S.
OBJ.
CUBE
6
5
UEP I2A
USB
RS2 32C
LCN T
EPI brightness switch
EPI
DIA
21

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents