Polarizer Slider (For The Episcopic Illumination) - Nikon Eclipse LV100DA-U Instructions Manual

Universal design microscope udn
Table of Contents

Advertisement

14

Polarizer Slider (for the Episcopic Illumination)

Polarizer sliders
The three polarizer sliders as shown below can be used for the product.
Polarizer
LV-PO polarizer
LV-UPO polarizer
LV-UVPO UV polarizer
• LV-PO polarizer
For the simplified polarization microscopy under the episcopic illumination, use the polarizer with
the analyzer slider. For the sensitive color polarization microscopy under the episcopic illumination,
use the polarizer with the analyzer slider and the lambda plate slider.
When the motorized universal quintuple nosepiece (LV-NU5A or LV-NU5AC) is attached, the
prism slide method of the DIC microscopy under the episcopic illumination can be performed by
using the polarizer with the analyzer slider and the DIC slider.
• LV-UPO polarizer (equipped with the 1/4 lambda plate)
The LV-UPO is used to perform the Senarmont method of DIC microscopy when used with the
analyzer and the DIC slider. The contrast of the DIC contrast image can be adjusted with the
polarizer orientation.
• LV-UVPO UV polarizer
The polarizer is used to change the excitation light to a linearly polarized light for the epi-fl
microscopy under the UV excitation light. As the polarizer deteriorates over time, change it if
necessary.
56
Application
Polarization microscopy under the episcopic
illumination (simplified/sensitive color)
DIC microscopy under the episcopic
illumination (prism slide method)
DIC microscopy under the episcopic
illumination (Senarmont method)
Epi-fl microscopy
Remark
Equipped with the 1/4
lambda plate