Animal Weighing V - Sartorius LP Models Operating Instructions Manual

Master pro series electronic precision balances
Table of Contents

Advertisement

Animal Weighing V

Purpose
Use this program to determine the
weights of unstable samples (e.g.,
live animals) or to determine weights
under unstable ambient conditions.
In this program, the balance
calculates the weight as the average
of a defined number of individual
weighing operations. These
weighing operations are also known
as "subweighing operations."
42
Available Features
– Animal weighing started manually
or automatically
– Optional balance configuration
in the Setup menu for automati-
cally initializing this application
when you turn on the balance
– Minimum load threshold for
starting animal weighing:
– 100 display increments for
automatic start
– 50 display increments for
manual start
– Start range: Automatic start of
animal weighing operation when
three successive subweights
lie within a user-defined tolerance
range (calm = 2%, normal = 5%,
active = 10%)
– Number of weighing
operations for calculation of an
average mdef can be set before
the beginning of each animal
weighing operation
– Arithmetic average displayed as
a result in the pre-set weight unit
(identified by the G symbol)
– Optional multiplication
of the arithmetic average by
a user-defined factor mul.
A circle "o" is displayed as
weight unit and mul=xxxx is
shown in the text line
– Toggling between weighed and
calculated results by pressing w
– Automatic output via the
interface port:
– Number of weighing
operations mDef
– Multiplication factor Mul
– Automatic output of results via the
interface port:
– Weighing result x-Net
– Calculated result x-Res
– Stop limit: Unload threshold
(50 display increments)
– Return to weighing mode
by unloading the balance;
i.e., when the load
is below the stop threshold
Factory Settings
Auto-start application (automatic
initialization with automatic start of
animal weighing): off (3 10 2)
Animal activity (3 subweights
lie within a preset range; i.e., animal
is calm, normal or active):
normal (3 7 2)
Start animal weighing:
automatic (3 8 2)
Automatic printout of number
of weighing operations used in
averaging and of the
calculation factors: off (7 1 1)
Print animal weights (automatic
printout of weighed or calculated
result):
On: animal wt. (3 9 2)

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents