Preparation
To calculate a piece count, the
average weight of one piece must
be known. This average piece
weight can be entered into the
Counting program in one of three
ways:
– The last reference sample
quantity entered is loaded and
displayed when you turn on the
balance. Place the same number
of parts on the balance and
initialize the counting program;
– With automatic initialization
switched on, the balance goes
into the "counting" mode when
you turn it on and loads the
last average piece weight and
corresponding reference
sample quantity that were entered
or calculated;
– Enter the average piece
weight using the numeric keys
and store it.
Reference Sample Updating
When the updat soft key is
displayed during counting, this
means you can have the average
piece weight updated (while
the piece count is displayed in the
measured value line).
The updat soft key is displayed
when:
– the balance has reached stability
– the current piece count is less than
double the original piece count
– the current piece count is less
than 100
– the internally calculated piece
count (e.g., 17.24 pcs) differs from
the nearest whole number (here:
17 pcs) by less than 0.3
Reference sample updating can be
repeated several times with an
approximately doubled piece count.
34
To perform reference sample
updating: Press v (updat
soft key)
Turn on the balance:
Press e
> All display segments light up
briefly
Select the Counting program
in the Setup menu: Press M
Select menu:
Press the menu soft key (w)
Set parameter 2 1 4: See
"Configuring the Balance"
Exit the Setup menu:
Press M
Setting Parameters for the
Counting Application
Access the Setup menu:
Select Menu
Set parameters for:
– Storage accuracy:
3 5 Count/pct. param.
– Automatic initialization:
3 10 auto-start app.
– Automatic output of
parameters to interface port:
7 1 Print app. param.
See "Configuring the Balance"
Exit the Setup menu:
Press M
Additional Functions
In addition to the functions for:
– alphanumeric input,
– taring
(not during alphanumeric input),
– printing,
you can also access the following
functions from this application:
– calibration
(not during alphanumeric input),
– setup,
– turning off the balance.
Calibration/Adjustment
Press q
> See "Calibration/Adjustment"
for further instructions
Setup (setting parameters)
Press M
> See "Configuring the Balance"
for further instructions
Turning Off the Balance
Press e
> The balance shuts off
> The display goes blank