Practical Examples
Example P1: Weighing in Percent with Reference Weight Taken from Weight on Balance
Step
1. Select the weighing-in-percent
application in the Setup menu
2. Prepare a container for the parts
3. Tare the balance
4. Place the reference weight
on the balance
(here: 1821.48 g =100%)
5. Initialize the balance
6. Unload the balance
7. Determine the percentage
of an unknown weight
8. If desired, print percentage
(here: 98.37%)
38
Key (or instruction)
see "Preparation"
Place the empty container
on the balance
t
Place weight equal
to reference weight in
the container
start soft key v
Remove reference weight
from the container
Place sample to
be measured in the container
p
Display/Output
wxx%=182148
wxx%=182148
wxx%=182148
pRef
+
100 %
Wxx%
+
1821.48 g
Prc
+
98.37 %
.
.
.