External Calibration/Adjustment*
with a User-Defined Weight
First set either 1 9 3 or selection
mode (1 9 12) in the menu.
You can define a weight for
calibration/adjustment.
External calibration/adjustment must
be performed with weights that are
traceable to a national standard and
Define the Calibration Weight
Step
1. Select the Setup menu
2. Access the Input Mode
3. Select input for calibration
weight (currently 3000.00 g)
4. Enter calibration weight
(e.g., 4000.00 g)
5. Save weight
6. Exit the Setup menu
* = for verified balances of accuracy class K, only external calibration is possible
that have error limits which are at
least 1/3 of the required tolerance
of the display accuracy.
The balance has a factory-set weight
value (see "Specifications").
Key (or instruction)
M
Input soft key v
Y Y Y
4 0 0 0 . 0
0
... weight soft key v
M
Operating the Balance
To reset a user-defined calibration/
adjustment weight to the original
factory setting, enter the factory-set
weight value manually (see
"Specifications") or set menu code
1 9 1 in the Setup menu.
Display/Output
info--menu--input
27