7.12.3 Replacing a defective locking spring

CAUTION!
Danger of tipping over due to defective locking springs. Defective locking springs can lead
to loss or failure of the anti-tipping mechanism!
· Check carefully that the locking springs are intact.
Requirements:
· A rod or Phillips screwdriver
· If necessary press the defective locking spring
together.
· Pull the anti-tipping mechanism (1) rearward
out of the support tube (3).
· Remove the defective locking spring.
If it is not possible to remove the defective
locking spring, push the defective locking
spring (2) into the support tube using the rod
or Phillips screwdriver.
· Insert a new locking spring (4) into the anti-
tipping mechanism up to the second hole.
· Insert the anti-tipping mechanism into the
support tube again as far as it will go. To do
this, press the locking spring together slightly.
· Make sure that the locking spring latches into
the holes provided for this purpose in the
support tube.
®