SERVICE MANUAL
7.5
Operating hours counter
7.5.1
Replacing the operating hours counter
Requirements:
· Philipps screwdriver, size 2
· 10 mm open-end wrench
· Switch the electronics OFF.
· Remove the legrests.
· Release the seat (1) and tilt forward.
· Disconnect the remote bus cable from the
electronics module or ACT.
· Lift the seat up out of its forward mounting
and remove completely. The seat has a
heavy weight. Use proper lifting techniques.
The operating hours counter is attached to the
frame with a Velcro strip.
· Disconnect the cable of the operating hours
counter from the drive motor.
· Pull the operating hours counter off of the
frame.
· Turn the operating hours counter over.
· Loosen and remove the four Philipps head
screws of the casing.
· Disconnect the plugs (+) & (-).
· Remove the nut (4) including the locking
washer.
· Pull the operating hours counter (6) forwards
out of the casing (1). Pay attention to the
mounting bracket (5) and the retaining frame
(7).
· Replace the operating hours counter.
58
®
®
Invacare
- Bora
/Spectra
®
XTR Series