Vr2 Remote: Error Codes And Diagnostic Codes - Invacare Bora XTR Series Service Manual

Hide thumbs Also See for Bora XTR Series:
Table of Contents

Advertisement

®
®
Invacare
- Bora
/Spectra
6.3

VR2 remote: Error codes and diagnostic codes

Evaluate the cause using the following blink codes. The following figure shows which LEDs are
located on the remote.
1) Battery display
2) Profile indicator
3) Adjusting motors
Error code
1 LED battery display
2 LED battery display
3 LED battery display
4 LED battery display
5 LED battery display
6 LED battery display
7 LED battery display
8 LED battery display
®
XTR Series
Possible Cause
Batteries
discharged
Error at left-hand
motor
Connection
loose/defective
or motor
defective
Short-circuit in
left-hand motor
Error at right-
hand motor
Connection
loose/defective
or motor
defective
Short-circuit in
right-hand motor
The mobility
device has been
blocked by an
external signal,
for example
because the
charger is
connected.
Fault on drive
lever
Fault in the
electronics
Solution
Charge battery
Check cable to batteries
Check connection plug, check
motor
Check connection plug, check
motor
Check connection plug, check
motor
Check connection plug, check
motor
Remove battery charger
Put the drive lever in a central
position before switching the
remote on.
Check cable
SERVICE MANUAL
Documentation
See chapter 7.9.
See chapters 7.9
and 7.2
See chapters 7.9
and 7.2
See chapters 7.9
and 7.2
See chapters 7.9
and 7.2
See chapter 7.9.
33