SERVICE MANUAL
7.5.2

Replacing the connecting cable of the operating hours counter

Requirements:
· Philipps screwdriver, size 2
· Side cutters
· Cable ties
· Switch the electronics OFF.
· Remove the legrests.
· Release the seat (1) and tilt forward.
· Disconnect the remote bus cable from the
electronics module or ACT.
· Lift the seat up out of its forward mounting
and remove completely. The seat has a
heavy weight. Use proper lifting techniques.
The operating hours counter is attached to the
frame with a Velcro strip.
· Disconnect the cable of the operating hours
counter from the drive motor.
· Pull the operating hours counter off of the
frame.
· Turn the operating hours counter over.
· Loosen and remove the four Philipps head
screws of the casing.
· Disconnect the plugs (+) & (-).
· Remove the cable ties (4).
· Carefully pull the connecting cable (2) of the
operating hours counter out of the casing (1).
· Replace the cable of the operating hours
counter.