7.19.2 Replacing the lifter actuator

Requirements:
· Screwdriver flat, width 6 mm
· Large screwdriver
· Flat screwdriver
·
Remove legrests.
·
Run the lifter up to if possible.
·
Switch the electronics OFF on the remote.
·
Secure the seat.
·
Pull the plug from the actuator.
·
Bottom fixing bolt (2) secured with snap
rings – Remove a snap ring using a
screwdriver and remove the bolt.
·
Top fixing bolt (3) with slot –
Loosen the fixing bolt with a screwdriver.
Remove the bolt.
·
Replace actuator.
·
Reassembly of the actuator takes place in
the reverse order.
·
Connect the plug of the actuator cable to the
actuator.
·
To complete, check all seat tilting function.
If the lifter cannot be raised, a cover can be removed from the upper rear side of the module. This
way it is possible to reach the lower fixation bolt of the actuator.
®