Error code
10 LED battery display
7 LED Battery display
plus
5 LED
profile indicator
Actuator LEDs blinking Actuator error
®
XTR Series
Possible Cause
Power surges in
the control box,
e.g. due to bad
connection with
batteries.
Compatibility
problems
between modules
Solution
Check cable to batteries
Check cable to remote
Replace remote
If more than one actuator is
fitted, locate the defective
actuator.
Check cable to actuator
SERVICE MANUAL
Documentation
See chapter 7.9.
See chapter 7.9.
See chapter 7.9.