Point to Settings and press the joystick to display a Line settings dialog box where
you can change the settings for the line.
10588803;a2
Figure 10.32 Line dialog box
Figure 10.33 Explanations of the Line dialog box
Label
Local
Emissivity
Publ. No. 1558240 Rev. a156 – ENGLISH (EN) – March 1, 2006
Explanation
Point to Avg and press the joystick to display the average tempera-
ture along the line.
See below.
Value
On
Off
User-defined
(0.01–1.00)
Comments
Select On to set the emissivity, the reflected tem-
perature, and the distance for this line only.
Selecting On will also assign an asterisk to the
measurement marker's label.
You can set the Emissivity if Local is enabled. If
not, this option will be shaded.
➲ If you enter an emissivity value less than 0.30
the emissivity box will begin flashing to remind you
that this value is unusually low.