Replacing Excitation And Barrier Filters; Filter Cube Internal Spacers - Nikon Eclipse 50i Instructions Manual

Hide thumbs Also See for Eclipse 50i:
Table of Contents

Advertisement

Replacing excitation and barrier filters

3.17.3
Excitation and barrier filters can be removed from
the fluorescent cube and replaced. (Dichroic
mirrors cannot be dismounted from the fluorescent
cube.) Excitation filters are screw-in filters.
Barrier filters are slide-in filters. Align the
projection on the barrier filter with the groove on
the fluorescent cube and turn clockwise
approximately 30 degrees to secure in place.

Filter cube internal spacers

3.17.4
The filter cubes listed below cannot be inserted
directly into the fluorescence attachment. Instead,
as described below, an internal spacer must first
be removed or reversed.
(1) Place the filter cube on the work surface with
the excitation filter facing up.
(2) Unscrew the ring retaining the excitation
filter.
(Be careful to avoid dropping the filter.)
(3) Remove the spacer inside the ring.
(4) Remove or reverse the spacer as appropriate
for the particular filter cube type. The actions
suitable for various filter cubes are given
below.
(5) Reattach the ring.
- Filter cubes requiring spacer removal
• UV-2A
• UV-2B
- Filter cubes requiring spacer reversal
• DAPI
• FITC
• GFP-L
• GFP-B
• TRITC
• Tx-Red
Chapter 3 Individual Operations
3.17 Selecting Fluorescent Filters
57
Turn about 30° to secure in place.
Excitation
filter
Filter cube
Spacer
Ring
Barrier filter
3

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Eclipse 55i

Table of Contents