Animas OneTouch Ping Owner's Booklet page 94

Glucose management system
Hide thumbs Also See for OneTouch Ping:
Table of Contents

Advertisement

CHAPTER 13 - SAFETY SYSTEM AND ALARMS
Alerts, Warnings and Alarms
Warning: Exceeds Max TDD
Cause
Bolus delivery exceeds user-set maximum.
Effect
All deliveries stop until action is taken. Any Combo
Bolus or Temp Basal is temporarily suspended.
Message
Displayed when manually awakened until confirmed.
Press
Setup Advanced menu. If the Warning is not confirmed
Action
by the time your pump clock passes midnight, the
message will continue to be displayed, but any Combo
Bolus or Temp Basal that is currently suspended will
resume.
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.
Warning: Exceeds Max Basal
Cause
Basal delivery rate (or Temp Basal delivery) exceeds
user-set maximum.
Effect
Basal delivery stops.
Message
Displayed when awakened (by basal delivery attempt
every 3 min. or manually) until confirmed.
Action
Press
the Setup Advanced menu (or reprogram Temp Basal).
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.
to confirm. Reprogram max TDD amount in the
to confirm. Reprogram Max Basal amount in
80
Warning
Exceeds
max TDD
XXX U.
No delivery.
Confirm
Warning
Exceeds
max basal
XXX U/Hr.
No basal
delivery.
Confirm

Advertisement