Animas OneTouch Ping Owner's Booklet page 198

Glucose management system
Hide thumbs Also See for OneTouch Ping:
Table of Contents

Advertisement

CHAPTER 2 - ESTABLISHING COMMUNICATION BETWEEN YOUR METER REMOTE AND YOUR PUMP
After your devices are paired, the ezCarb and ezBG Bolus screens on your
meter remote display will retrieve the bolus calculator values (settings) that
are set and saved on your pump. You will not have access to the Calculator
Set-up option on your meter remote. You will be reminded on your meter
remote display that the bolus calculator settings from Calculator Set-up on
your meter remote have been replaced.
Press
on your meter remote to confirm the Notification. You will go to the
Pump Home screen on your meter remote display (see Chapter 3 in Section III, pages 187–188).
NOTE: Your meter remote and pump must use the same unit of measure (mg/dL) for BG measurements
or the devices cannot be paired. The BG unit of measure for both devices is set at the factory and
cannot be changed.
If the BG unit of measure is not the same on both devices, you will be
notified on your meter remote display during the pairing procedure.
Contact Customer Service at 1 877 937-7867 for instructions on replacing
your meter remote or pump with one that has the correct glucose unit of
measure.
This Notification screen will be followed by a Warning screen on your meter
remote display indicating that the pairing procedure has failed. Press
your meter remote to confirm the Warning. You will not be able to use your
meter remote to access pump functions unless both devices have the same
glucose unit of measure (mg/dL).
Performing a new pairing with a replacement meter remote or pump
If you obtain a replacement meter remote or pump, you will have to complete the pairing procedure
again so that the new device is recognized. Any new pairing will automatically cancel the previous
pairing.
1
Reactivate the pairing feature on your pump
Go to the SETUP ADV 10 screen on your pump. Press
the "Search" field is highlighted on your pump display. Press
so that the highlight is flashing. Press
to "ON" and press
to reactivate the pairing feature on your pump.
on your pump so that
on your pump
or
on your pump to change "OFF"
180
Notification
Stored meter calculator
settings will be
overwritten by settings
from pump.
Confirm
Meter remote display
Notification
Pump and meter glucose
units do not match.
Contact customer service.
Confirm
Meter remote display
WARNING
on
Pairing procedure failed!
Confirm
Meter remote display
SETUP ADV 10
METER
RF
Search
Channel
[ABCD1234]
Home
Pump display
ON
OFF
7
Next

Advertisement

Table of Contents
loading

Table of Contents