CHAPTER 12 - CARE AND MAINTENANCE

CHAPTER 12 - CARE AND MAINTENANCE
The Vents
Your pump features a redundant Vent Safety System. Vents serve two purposes. First, they allow air to
enter and exit your pump so that pressure is equalized under a variety of environmental
circumstances, such as changes in altitude. Second, the vents are backed by a special membrane,
which keeps water from entering your pump.
Battery Cap with O-ring and Vent
Your battery cap contains an o-ring and vent. There is a tiny hole backed by a membrane, which allows
air to pass through but prevents water from entering. The o-ring helps to keep your pump waterproof. It
is recommended that you change the battery cap/vent twice a year. If you work in a dusty environment
such as a construction site, mill, cement factory, etc., or if you are a frequent swimmer, you should
change your battery cap every 3 months. You can call Animas
O-ring
WARNING: Under no circumstances should you introduce any kind of sharp object into the
vent openings to clean them. Doing so could compromise your pump's waterproof capabilities.
If at any time you suspect the vent opening is clogged, replace the battery cap.
Cleaning
CAUTION: Do Not use household or industrial cleaners, chemicals, solvents, bleach,
scouring pads or sharp instruments to clean your pump. Never put your pump in the dishwasher
or use very hot water to clean it. Use only a very mild detergent (for example a drop of liquid
soap in a glass of water) and a lint-free cloth.
Never put your pump in a microwave oven or baking oven to dry it. Use a soft towel.
Never clean the battery or insulin cartridge compartments.
General Wear and Tear
If you drop your pump or it has been hit against something hard, inspect it to be sure it is still working
properly. Check that the display screen is working and clear, that the cartridge cap, battery cap and
infusion set are properly in place. Check for leaks around the cartridge by wrapping a piece of tissue
around the connection area. Cracks, chips or damage to your pump may impact the battery contact
and/or the waterproof feature of your pump. Call our Customer Service representatives at
1 877 937-7867 if you identify or suspect your pump is damaged. They will help determine if your
pump should be replaced.
®
to order an extra battery cap.
73
Vent Hole