CHAPTER 13 - SAFETY SYSTEM AND ALARMS

Alerts, Warnings and Alarms
Alert: Temp Basal Minimum Rate
Cause
Negative Temp Basal activated.
Effect
Basal delivery will not go below 0.025U/Hr.
Message
Displayed once for 3 seconds.
Action
None required.
Beeps/Vib
User selected, one time. No progression.
Alert: Suspend (Temp Basal/Combo Bolus Canceled)
Cause
Pump suspended
Effect
Any active Temp Basal/Combo Bolus canceled.
Message
Displayed once for 3 seconds.
Action
None required.
Beeps/Vib
User selected, one time. No progression.
Alert: Low BG
Cause
BG entry below 70 mg/dL.
Effect
Requires user confirmation to continue.
Message
Displayed until confirmed or until pump goes to sleep.
Action
Press
Beeps/Vib
User selected, one time. No progression.
to confirm.
76
Alert
Temp Active
Minimum
Basal rate
limited to
0.025U/Hr
Alert
Pump Suspended
If active,
Temp Basal &
Combo Bolus
have been
canceled.
Alert
LOW BG
Treat low BG.
No bolus
recommended.
Monitor BG.
Confirm